peter.zhang.tup@gmail.com

Table of Contents

Application of 96-Well Plate in LC-MS/MS Analysis of Catecholamine Metabolites in Plasma Samples

 

Introduction:

The TUP WCX 96-well plate utilizes patented loading technology and high-quality packing materials, providing lower background interference and higher reproducibility for experiments. It is easy to operate and minimizes the risk of cross-contamination.

 

Sample preparation method:

 

Preparation of serum samples

Mix 375μL of serum with 10μL of 0.1% formic acid in acetonitrile, 10μL of internal standard, and 375μL of 25mM ammonium acetate solution. Vortex before injection.

Materials

WCX 96-well plate (TUP Catalog No.: H07-00502, 5mg/2ml).

Serum processing procedure
  • Activation:Activate with 200 μL methanol first, then activate with 200 μL 25 mM ammonium acetate, controlling the flow rate at 3-5 s/drop.
  • Loading:Load all pre-treated samples, apply positive pressure, control the flow rate at 3-5 s/drop, and discard.
  • Washing 1:Add 400 μL of 25 mM ammonium acetate to each well, apply positive pressure, control the flow rate at 3-5 s/drop, and discard.
  • Washing 2:Add 400 μL of acetonitrile to each well, apply positive pressure, control the flow rate at 3-5 s/drop, and discard.
  • Elution:Add 200 μL of 5% FA acetonitrile to each well, press down with positive pressure, control the flow rate at 3-5 s/drop, and collect.
  • Nitrogen blowing and reconstitution:The collected eluate was dried under nitrogen at 35 °C, and then reconstituted with 100 μL mobile phase (initial ratio), vortexed for 2 min, transferred to a sample vial, and detected by LC-MS.

 

LC-MS  conditions

 

HPLC conditions
  • Specialized column, 2.1 x 100 mm, 3 μm particle size, 120 Å pore size.
  • Column temperature: Set at 40℃.
  • Injection volume: Set at 10 μL.
  • Mobile phase conditions

A: 10 mM ammonium formate with 0.1% formic acid

B: 0.1% Formic acid, 10 mM Ammonium Formate in 90% Acetonitrile

Total time (min) Flow rate (μL/min) %A %B
0 500 0 100
0.1 500 10 90
0.6 500 10 90
1.2 500 40 60
3 500 40 60
3.01 500 0 100
6.5 500 0 100

 

MS conditions
  • Ionization mode: Electrospray ionization positive mode (ESI+)
  • Detection mode: Multiple Reaction Monitoring (MRM)
  • Electrospray voltage: 5500V
  • Ion source temperature: 500℃

 

ID Q1 Q2 DP CE
3-MT 1 168.1 119.1 50 26
3-MT 2 168.1 91.1 50 33
MN 1 180.1 148.1 90 25
MN 2 180.1 120.1 90 25
MN-d3 1 183.1 151.1 80 25
MN-d3 2 183.1 123.1 80 24
NMN 1 166.1 134.1 85 23
NMN 2 166.1 121.1 85 24
NMN-d3 1 169.1 137.1 80 25
NMN-d3 2 169.1 109.1 45 19
3-MT-d4 1 172.1 123.1 55 26
3-MT-d4 2 172.1 95.1 55 35
E1 184.1 166.1 60 12
E2 184.1 107.1 80 26
E-d6 190.1 172.1 60 14
NE 1 152 107 60 24
NE 2 152 135 60 17
NE-d6 1 158.2 111.1 76 25
NE-d6 2 158.2 140 76 20
DA 1 154.2 137.1 40 15
DA 2 154.2 97.1 90 26
DA-d4 158.2 141.1 40 16

 

Results and Discussion

 

Linear Range and Sensitivity

A calibration curve was prepared using 2% BSA as the matrix with six different concentrations (10, 20, 50, 100, 200, and 400 pg/ml). The samples were processed using the aforementioned pre-treatment method and subjected to analysis. The results obtained are presented in the table below.

 

ID Regression equation Correlation coefficient
MN y=3.52x+0.167 y=3.52x+0.167
NMN y=8.77x+0.552 y=8.77x+0.552
3MT y=18.13x+0.902 y=18.13x+0.902
NE y=0.00612x+0.00721 y=0.00612x+0.00721
E y=0.00316x+0.00164 y=0.00316x+0.00164
DA y=0.00543x+0.00211 y=0.00543x+0.00211

 

Recovery results
ID Spiked concentration

pg/ml

Average recovery rate

(n=3)

RSD

(n=3)

MN 20 20 20
99.1% 99.1% 99.1%
NMN 20 20 20
98.7% 98.7% 98.7%
3MT 20 20 20
99.9% 99.9% 99.9%
NE 20 20 20
99.3% 99.3% 99.3%
E 20 20 20
101.8% 101.8% 101.8%
DA 20 20 20
99.2% 99.2% 99.2%